| Measuring Geometrics |
- 45° illumination
- 15°, 25°, 45°, 75°, 110° aspecular viewing
- Angular Accuracy ±0.15°
- Fibre optic pick-up, coupled with DRS technology
|
|
Measurement Area |
0.5" (12mm) diameter |
|
Light Source |
Gas-filled tungsten lamp, color corrected to approximately 4000°K |
|
Illuminant Types |
C, D50, D65,D75, A, F2, F7, F11 and F12 |
|
Standard Observers |
2° and 10° |
| Receiver |
Blue-enhanced silicon photodiodes |
|
Spectral Range |
400 - 700nm |
|
Spectral Interval |
28 band spectral measurement:
10nm interval from 400nm - 640nm
20nm interval from 640nm - 700nm
15nm bandwidths |
|
Spectral Data Output |
Spectral reflectance values are available for output from the RS232 port for 5 angles at 10nm intervals from 400nm700nm |
|
Measurement Range |
0 to 400% reflectance |
|
Measuring Time |
Approximately 2 seconds |
|
Inter-Instrument Agreement |
- 0.18 DE* average on reference BCRA tile set
- 0.35 DE* maximum on any chromatic tile
- 0.15 DE* maximum on any grey tile
|
|
Short-Term Repeatability |
0.10 DE*abon white ceramic (standard deviation) |
|
Lamp Life |
Approximately 500,000 measurements |
|
Power Supply |
Six rechargeable AA Ni-metal hydride batteries included - removable battery pack; 7.2 VDC rated @ 1400 mAh |
|
AC Adapter Requirements |
- MA68: 90 - 130VAC, 50 - 60Hz, 15W max
- MA68X:180 - 260VAC, 50 - 60Hz, 15W max
- 12 VDC @ 700mA
|
|
Charge Time |
In Instrument: 4 hours (50%), 16 hours (100%) |
|
Measurements per Charge |
1,000 5-angle measurements |
|
Data Storage (Five Angles) |
200 Standards, 850 Samples |
|
Data Interface |
Patented bi-directional RS232, 300 - 19,200 baud |
|
Display |
4-row by 20 character supertwist dot matrix LCD |
|
Operating Temperature Range |
50° to 104°F (10 to 40°C)
85% Relative Humidity Maximum (non-condensing) |
|
Storage Temperature Range |
-4° to 122°F (-20° to 50°C) |
|
Weight |
3.2lbs (1.4kg) |
|
Dimensions |
4.56" x 3.0" x 8.85" (116 x 76 x 225mm) |
|
Supplied with |
Calibration Standards, Operating Manual, AC Adapter and Carrying Case |