An enhancement
to the 7000/1 Setaflash Series 7 with the additional feature of
electronic Peltier cells for rapid heating and cooling over its
operating range of -30°C to 100°C (-22°F to 212°F).
This allows
much greater flexibility, without the lengthy cool down experienced
with other flash point testers. A supply of cooling water is required
and to achieve -30°C (-22°F) the supply must be at least
0.6 litres/minute at 5°C (41°F) or lower.
Each instrument
is supplied with a -30°C to 100°C (-22°F to 212°F)
secondary reference thermometer.
Also available
is the Elcometer 7000/3 Setaflash Series 7 Plus 'Special', which
is the same as the 7000/1 Setaflash but is fitted with a cup suitable
for testing corrosive samples.
See also Elcometer 7000/1 Setaflash 'Series
7'