A simple to
use, continuous humidity chamber to meet the test condition required
by ISO 6270. The Elcometer 1200/4 is of welded polypropylene tank
construction. The PVC cover has two sides, each at an angle of 15
degrees to the horizontal.
The Unit is
switched on and water is fed in, through the inlet pipe, under pressure
(1-5 bar). Once the water is at the required level, the unit begins
to heat and maintain the water to the set temperature (40°C
± 2°C / 104ºF ± 4ºF, and not less than
35°C / 95ºF for ISO 6270).
Up to 20 test
panels, 150 x 100 mm (5.9 x 3.9"), are then inserted into the
spaces in the PVC cover.
Note: To meet
the ISO 6270 Standard, all 20 must be filled.